The FD-800 Infrared Moisture Determination Balance employs a dual temperature sensing method to achieve highly precise moisture measurements. In addition to the thermistor that measures the environment temperature within the drying chamber, it is mounted with a radiation thermometer that measures the sample temperature directly without contact.

SpecificationsFD-800
Measurement methodDetection of weight loss by heating & drying
Sample mass0.1-120g (optional weight sampling)
Measurement subjectMoisture / Solid content / Weight
Measurement range0-100% (wet base, solids), 0-500% (dry base)
Repeatability
(standard deviation)
Sample with a weight of 5g or higher : 0.05%
Sample with a weight of 10g or higher : 0.02%
ResolutionMoisture percentage 0.01% / 0.1%; Weight 1 mg Temperature setting range for thermistor (T1): 30 - 180°C (1°C interval); for radiation thermometer (T2) 30 - 250°C (1°C interval)
Measurement modesAutomatic stop mode, Timer stop mode, Rapid drying mode, Equilibrium drying mode, Step drying mode, Predictive (comparative) measurement mode
External outputRS-232C interface
Environmental conditionsTemperature 5~40°C , Humidity : max of 85%RH Pollution degree 2 , Altitude up to 2000m
Sample panSUS (diameter 130 mm, depth 13 mm)
Heat sourceMid-wave infrared quartz heater (200Wx2)
Power sourceAC100~120V / 220~240V (50/60Hz)
Power consumptionMax 900W
Dimensions & weight220(W) x 415(D) x 190(H) , Net 4.5Kg , Shipment 10.0Kg
AccessoriesSpare sample pan, Aluminum sheets, etc.
Options

Printer Set (VZ-330, Printer Cable: VZC-14) 10x Printer paper rolls, 500 x Aluminum sheets Radiation thermometer calibration set GF-200 (digital thermometer, standard heating sample), FD- 800 Data logger software FDL-01 (RS-232C cable, USB-RS232C conversion cable), Windshield with deodorizer case FW-100

 

Product Brochure: Model FD-800 - 381 kB


Product Manual: Model FD-800 - 5464 kB